Description and Specification for Lab Tenders
Astra Scientific Free and Force Convection Apparatus.
The Offers basic experiments for targeted teaching on the topic of free
and forced convection on various heating elements. At the heart of the
experimental unit is a vertical air duct into which various heating
elements are inserted. An axial fan is located on top of the air duct.
The fan draws in ambient air and guides it through the air duct. The air
flows past a heating element and absorbs heat. Four heating elements
with different geometries are available to be selected. In order to
investigate free convection, two of the four heating elements can be
operated outside of the air duct. The heating elements are designed in
such a way to release heat only at their surface. The compact design
ensures rapid heating and a short time for experiments.
The experimental unit is equipped with temperature sensors at the inlet
and outlet of the air duct. The air velocity is measured to determine
the airflow rate. Heating power and flow rate are adjusted and displayed
via the software.
The microprocessor-based instrumentation is well protected in the
housing. The software consists of a software for system operation and
for data acquisition and an educational software. is connected to the PC
via USB.
FEATURES:
• Free and forced convection
• Calculation of convective heat transfer at different geometries
o Flat plate
o Cylinder
o Tube bundle
• Experimental determination of the Nusselt number
• Calculation of typical characteristic variables of heat transfer
o Nusselt number
o Reynolds number
• Investigation of the relationship between flow formation and heat transfer during experiments
• Description of transient heating process
• Functions of the GUNTsoftware: educational software, data acquisition, system operation
SPECIFICATION:
• Air duct
Flow cross-section: 120x120mm
Height: approx. 0,6m
Heating elements, temperature limitation: 90°C
• Tube bundle
Number of tubes: 23
One tube in variable position is heated
Heating power: 20W
Heat transfer area: 0,001m2
• Cylinder with an even temperature at the surface
Heating power: 20W
Heat transfer area: 0,0112m2
• Plate
Heating power: 40W
Heat transfer area: 2x 0,01m2
• Cylinder with heating foil to investigate the local heat transfer
Heating power: 40W
Heat transfer area: 0,0112m2
• Axial fan
Max. Flow rate: 500m3/h
Max. Pressure difference: approx. 950Pa
Power consumption: 90W
• Measuring ranges
Air velocity: 0…10m/s
Temperature: 4x 0…325°C
Heating power: 0…50W
• Required for Operation
230V, 50Hz, 1 phase
230V, 60Hz, 1 phase; 120V, 60Hz, 1 phase