Description and Specification for Lab Tenders
Astra Scientific Computerized Cooling Tower Test Bench.
The examines the main components and principle of a wet cooling tower
with forced ventilation. Water is heated in a tank and transported by a
pump to an atomizer. The atomizer sprays the water to be cooled over the
wet deck surface. The water trickles from the top to the bottom along
the wet deck surface whilst air flows from the bottom to the top. The
heat is transferred directly from the water to the air by convection and
evaporation. The evaporated water volume is recorded. The airflow is
generated by a fan and adjusted using a throttle valve.
The cooling column is transparent allowing clear observation of the wet
deck surface and the trickling water. Interchangeable cooling columns
enable comparative studies. Software for data acquisition via USB.
All important process parameters are recorded (volumetric air flow rate,
temperatures of air and water, air humidity, water flow rate). The
measured values can be read on digital displays. At the same time, the
measured values can also be transmitted directly to a PC via USB. The
data acquisition software is included. The changes of state of the air
are represented in an h-x diagram.
FEATURES:
• Thermodynamic principles of the wet cooling tower
• Changes of state of the air in the h-x diagram
• Determination of the cooling capacity
• Energy balances
• Calculation of process parameters, such as maximum cooling distance, cooling zone width etc.
SPECIFICATION:
• Cooling column
Specific surface of the wet deck surface: 110m2/m3
Cross-section: 150x150mm
Volumetric airflow measurement via orifice: Ø 80mm
• Heater, adjustable in three stages:
500W
1000W
1500W
Thermostat: switches off at 50°C
• Fan
Power consumption: 250W
Max. Pressure difference: 430Pa
Max. Volumetric flow rate: 13m3/min
• Pump
Max. Head: 70m
Max. Flow rate: 100L/h
Tank for additional water: 4,2L
• Measuring ranges
Differential pressure: 0…1000Pa (air)
Flow rate: 12…360L/h (water)
Temperature: 2x 0…50°C, 3x 0…100°C
Rel. humidity: 10…100%
• Required for operation
230V, 50Hz, 1 phase
230V, 60Hz, 1 phase