Description and Specification for Lab Tenders
Astra Scientific Copper Strip Corrosion Test Apparatus.
Standard: ASTM D130; IP 154 FSPT DT-28-65; ISO 2160; DIN 51759; FTM 791-5325; NF M 07-015
Test Method: The Copper Strip Tarnish Test assesses the relative degree
of corrosivity of petroleum products, including aviation fuels,
automotive gasoline, natural gasoline, solvents, kerosene, diesel fuel,
distillate fuel oil, lubricating oil and other products. A polished
copper strip is immersed in 30mL of sample at elevated temperature.
After the test period, the strip is examined for evidence of corrosion
and a classification number from 1-4 is assigned based on a comparison
with the ASTM Copper Strip Corrosion Standards. For aviation fuels and
natural gasoline the sample tube is placed inside a stainless steel bomb
during testing.
TEST BOMB BATHS: Thermostatically controlled water bath immerses Copper
Strip Corrosion Test Bombs at the required depth per ASTM
specifications. Use for testing aviation gasoline, aviation turbine fuel
and natural gasoline. Fully insulated, double-wall stainless steel
construction. Soxhlet reflux condenser and constant water level device
maintain proper working depth. Choice of four-bomb and eight-bomb
models. Optional removable test tube rack converts four-bomb model for
testing of products not requiring corrosion bomb.
Specifications:
Conforms to the specifications of: ASTM D130; IP 154 FSPT DT-28-65; ISO 2160; DIN 51759; FTM 791-5325; NF M 07-015
Testing Capacity - K25310/K25319: four (4) copper strip corrosion test
bombs: eight (8) copper strip corrosion test bombs *or sixteen (16) test
tubes with optional test rack
Maximum Temperature: 221°F (105°C)
Temperature Control Stability: ±1°F (± 0.5°C)
Heater Range: 0-750W
Bath Medium: 5 gal (18.9L) water
Temperature Control: Analog
Electrical Requirements: 115V 60Hz, Single Phase, 7.5A 7nbsp; ; 220-240V 50/60Hz, Single Phase, 4A
Specifications:
Conforms to the specifications of: ASTM D130, D6074, D6158; FSPT
DT-28-65; IP 154; ISO 2160; DIN 51759; FTM 791-5325; NF M 07-015
Capacity: 17 test tubes
Maximum Temperature: 190°C (374°F)
Temperature Control Stability: ±1°C (±2°F)
Heater Range: 0-750W
Bath Medium: 5 gal (18.9L) water or high temperature heater transfer fluid
Temperature Control : Digital
Electrical Requirements: 115V 60Hz, Single Phase, 7.5A ; 220-240V 50/60Hz, Single Phase, 4A